Bed-exit alert—
Confirmed state
⬚
Vacant
☾
Resting
☀
Active
for 00:00
●—
This session
Bed exits
0
Last bed exit
—
Resting time
00:00
Active time
00:00
Block probability P(state)
block #0raw: — · q —
Live values
present
—
ρ (jitter ratio, zone-max)
—
a (activity, P90 excess)
—
bed zones B
—
Static — per zone
Dynamic — per zone
State timeline
Raw (per block)
Confirmed
Vacant
Resting
Active
Event log
No events yet
Calibration by example
Record three examples in order. The system learns each state automatically — no numbers to tune.
Record in order: ① empty bed → ② lie still → ③ normal activity. Bed zones and per-zone noise floors are learned automatically from ① and ②.
Settings
Time base, confirmation counts, and the bed-exit alert.
Bed-exit alert
Alert when the bed becomes vacant after being occupied.
Alert after
Grace period before the alert fires. "Immediately" is recommended for fall-risk patients.
Alert sound
Night hours only
Only alert between these hours (e.g. 21:00–06:00).
—
Time base T
Length of one block. The demo runs on 5 s.
Confirmation per transition
How many consecutive blocks confirm each from→to change. Resting→Vacant defaults higher: breathing pauses in real data look briefly like an empty bed.
Device
Location
Advanced
Internal knobs. Field staff normally never touch these.
v1.1: per-zone noise floors, zone-max ρ in ln(1+ρ) space, activity from dynamic P90 excess over per-zone floor.